76
Parameter/ Return
parameter
<NR2>
0.000 ~ 109.9
Example
MANU:ACW:CLOS 20.0
Sets the ACW LO SET current to 20 mA.
4.8.8 MANU:ACW:TTIMe
Set
Query
Description
Sets or returns the ACW test time in seconds. The test must
first be in ACW mode before this command can be used.
Note: A “TIME ERR” will result if the Ramp Time + Test Time
is ≥ 240 seconds when the HI SET limit is over 80mA . This
applies to the ACW function only.
In special MANU mode, the TIMER can be turned off.
Syntax
Query Syntax
MANU:ACW:TTIMe {<NR2>|OFF}
MANU:ACW:TTIMe?
Parameter
<NR2>
OFF
0.5 ~ 999.9 seconds
TIMER OFF (special MANU mode).
Return parameter <NR2>
TIME OFF
0.5 ~ 999.9 seconds
TIMER is OFF (special MANU mode).
Example
MANU:ACW:TTIM 1
Sets the ACW test time to 1 second.
4.8.9 MANU:ACW:FREQuency
Set
Query
Description
Sets or returns the ACW test frequency in Hz. The test must
first be in ACW mode before this command can be used.
Syntax
Query Syntax
MANU:ACW:FREQuency {50|60}
MANU:ACW:FREQuency?
Parameter/ Return
parameter
50
50 Hz
60
60 Hz
Example
MANU:ACW:FREQ 50
Sets the ACW test frequency to 50Hz.
4.8.10 MANU:ACW:REF
Set
Query
Description
Sets or returns the ACW reference value in mA. The test must
first be in ACW mode before this command can be used.
The ACW reference value must be less than the HI SET
value.
The ACW reference value must use the same range as the HI
SET value.
Syntax
Query Syntax
MANU:ACW:REF <NR2>
MANU:ACW:REF?
Parameter/ Return
parameter
<NR2>
0.000 ~ 109.9