75
4.8.5 MANU:ACW:VOLTage
Set
Query
Description
Sets or returns the ACW voltage in kV. The test must first be
in ACW mode before this command can be used.
Syntax
Query Syntax
MANU:ACW:VOLTage <NR2>
MANU:ACW:VOLTage?
Parameter/ Return
parameter
<NR2>
0.100 ~ 5.000 (kV)
Example
MANU:ACW:VOLT 1
Sets the ACW voltage to 1 kV.
4.8.6 MANU:ACW:CHISet
Set
Query
Description
Sets or returns the ACW HI SET current value in milliamps.
The test must first be in ACW mode before this command can
be used.
Syntax
Query Syntax
MANU:ACW:CHISet <NR2>
MANU:ACW:CHISet?
Parameter/ Return
parameter
<NR2>
0.001 ~ 110.0
Example
MANU:ACW:CHIS 10.0
Sets the ACW HI SET current to 10 mA.
4.8.7 MANU:ACW:CLOSet
Set
Query
Description
Sets or returns the ACW LO SET current value in milliamps.
The LO SET value must be less than the HI SET value. The
test must first be in ACW mode before this command can be
used.
The LO SET range must use the HI SET range. If all the digits
in the LO SET range are outside the HI SET range, an error
will be produced. All digits outside the HI SET range are
ignored and will not be used.
For example:
HI SET value: 12.34
LO SET value1: 0.005
error
LO SET value2: 0.053
no error
In the example above LO SET value1 will produce an error as
all digits are outside the range of HI SET. LO SET value2 will
not produce an error, but will return 0.05, not 0.053.
Syntax
Query Syntax
MANU:ACW:CLOSet<NR2>
MANU:ACW:CLOSet?