6.1 Fuse Replacement
Steps
1.
Turn the instrument off.
POWER
2.
Remove the power cord.
220
100
12
0
23
0
3.
Remove the fuse socket using
a flat screwdriver.
220
100
12
0
23
0
4.
Replace the fuse in the fuse
holder.
220
100
12
0
23
0
5.
Ensure the correct line voltage is lined up with the arrow
on the fuse holder. Insert the fuse socket.
220
100
12
0
23
0
220
100
12
0
23
0
Rating
STW-9900
STW-9800
100V/120V
T10A 250V
T5A 250V
220V/230V
T6.3A 250V
T2.5A 250V