
28
2.7 Temperature Measurements
The DL-2060 supports thermocouples and resistance temperature detector (RTD) types
of probes. For thermocouples, DL-2060 supports 7 types: E, J, K, N, R, S and T. Please
refer to
Table 3-1
for their temperature ranges. Be sure that the temperature function
is configured for the right sensor type before making measurements (Refer to 3.1.8 for
how to make the sensor configuration).
In general, RTDs have a better accuracy and long-term stability than thermocouples.
The default RTD type in DL-2060 is PT100 and the thermocouple is type K.
Table 3-1
Sensor Type
Temperature
Range(°C)
Temperature
Range(°F)
E
-270 to
1000
-518 to
1832
J
-210 to
1200
-140 to
2192
K
-270 to
1372
-518 to
2502
N
-270 to
1300
-518 to
2372
R
-50 to
1768
-122 to
3236
S
-50 to
1768
-122 to
3236
T
-270 to
400
-518 to
752
RTD (PT 100)
-200 to
850
-392 to
1562
2.7.1 Thermocouple Measurements
Connect the thermocouple adapter (TA-65 option: only for K-type) to the banana jacks
on the front panel. The difference between each type is subject to the patch thermal
leads.
※
Note:Only connection via the front panel can be used for temperature measurements.
How to measure thermocouple
①
Use terminals switch to select terminals on the front panel.
②
To measure thermocouple must be via thermocouple adapter and the low thermal
patch leads as shown in
Figure
3-12
. (When the mentioned modules are well
connected as the following figure, please go to the next step.)
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...