113
B.5 The Measurement Configuration Command
※
This mark means that it is available with firmware version 1.05 or upward.
[SENSe:]
FUNCtion “<
function
>”
Select a measurement function and enclose it in quotes in the command string (FUNC “VOLT:DC”).
Use one of the following strings.
VOLTage:DC VOLTage:AC VOLTage:DC:RATio
CURRent:DC CURRent:AC
RESistance (for 2-wire ohms) FRESistance (for 4-wire ohms)
FREQuenc PERiod CONTinuity
DIODe TCOuple TEMPerature
[SENSe:]FUNCtion?
Query the measurement function and return a quoted string.
[SENSe:]<
function
>:RANGe {<
range
>|MINimum|MAXimum}
Select a range for the selected function. For frequency and period measurements, range applies to
the signal’s input voltage, not its frequency (use FREQuency:VOLTage or PERiod:VOLTage). MIN
selects the lowest range for the selected function. MAX selects the highest range.
[SENSe:]<
function
>:RANGe? [MINimum|MAXimum]
Query the range for the selected function. For frequency and period, use FREQuency:VOLTage or
PERiod:VOLTage. Autorange thresholds: Down range at <10% of range; Up range at >120% of range.
[SENSe:]<
function
>:RANGe:AUTO {OFF|ON}
Disable or enable autorange for the selected function.
[SENSe:]<
function
>:RANGe:AUTO?
Query the autorange setting. Return “1” (ON) or “0” (OFF).
[SENSe:]<
function
>:RESolution {<
resolution
>|MINimum|MAXimum}
Set resolution for the selected function (not valid for frequency, period, or ratio). Specify the
resolution in the same units as the measurement function, not in number of digits. MIN selects
the smallest value accepted, which gives the most resolution. MAX selects the largest value
accepted which gives the least resolution.
[SENSe:]<
function
>:RESolution? [MINimum|MAXimum]
Query the resolution for the selected function.
[SENSe:]UNIT {Cel|Far|K}
Select units for temperature measurement. Cel means Celsius, Far means Fahrenheit,and K
means Kelvin.
Содержание DL-2060
Страница 1: ...B71 0110 41 INSTRUCTION MANUAL 6 1 2 DIGITAL MULTIMETER DL 2060 DL 2060G DL 2060VG DL 2060R DL 2060VR...
Страница 37: ...31 Figure 3 17 1 2 3 5 5 4...
Страница 135: ...129 C Tree Diagram Please refer to G Notice of Scanner Card if your firmware version is 1 05 or upward...
Страница 136: ...130...
Страница 137: ...131...
Страница 138: ...132 D DL 2060 s Dimension...
Страница 152: ...7F Towa Fudosan Shin Yokohama Bild 2 18 13 Shin Yokohama Kohoku ku Yokohama Kanagawa 222 0033 Japan http www texio co jp...