164
10-2. Vertical Accuracy Calibration
Panel Operation 1. Press the
Utility
key.
Utility
2. Press
System
from the bottom
menu.
3. Press
more 1 of 2
from the side
menu.
4. Press
Self Cal
on the side menu.
5. Press
Vertical
on the side menu.
6.
A message appears to “Now performing vertical
calibration…Set CAL to the channel, then press
the Vertical key”.
7. Connect the calibration signal from the rear
panel to the Channel 1 input with a BNC cable.
CH1
8. Press
Vertical
again after
connecting CAL to the channel 1
input.
The calibration for Channel 1 starts and ends
automatically, in less than 5 minutes. A
message is displayed when the calibration
procedure has ended.
9. Repeat the above step for Channel 2, 3 and 4
when prompted.
10. When the calibration for all channels has
completed, the display goes back to the default
state.