Low Voltage Test Procedure Example
14
SLUUC64 – June 2020
Copyright © 2020, Texas Instruments Incorporated
UCC5870QDWJEVM-026 Evaluation Module User’s Guide
10. Move to "Gate Voltage Monitor" section, disable the gate voltage monitor function or change gate
voltage monitor blanking time to 2500-ns and then change default state from "HiZ" to "PL", as this fault
may be triggered with 100-nF load and 5.1-
Ω
gate resistor if tGMBLK = 1000ns.
Figure 13. GUI Configuration 2 Disable Gate Voltage Monitor Function
11. Now move to Active State, setup PWM duty cycle for 50% and PWM frequency for 10kHz
Figure 14. GUI Active State Enable Continuous PWM