ADS1282-SP Setup and Quick Test
11
SBAU274 – July 2016
Copyright © 2016, Texas Instruments Incorporated
ADS1282-SP Evaluation Module (ADS1282EVM-CVAL)
Click the
Start Capture
button to plot data from the ADC. Click the
Frequency Domain
button to
measure the integrated signal-to-noise (
SNR
) in dBFS. In the following example, the 1
st
Nyquist integrated
SNR is 123.71 dBFs, very close to the typical datasheet value of 124 dBFs for this device configuration.
Figure 10. ADS1282-SP EVM GUI Data Capture Frequency Domain