3.11 DxSTRB Timing Adjustment
1. By default, the current firmware used by HSDC Pro sends a square wave instead of a pulse for the DxSTR
signals. The DAC requires a pulse for DxSTR that needs to be a width that is a multiple of 4 LVDS clock
cycles. To correct for this, do the following in HSDC Pro GUI:
a. Click on the
Instrument Options
tab located in the top left of the DAC main page.
b. Click on
IO Delay
.
Figure 3-8. IO Delay
c. Click the
Debug Features
button.
d. Enter x10000004 for the Reg Address and x8000 for the Data in the Write section as shown in
Write Registers
button. Enter x10000004 for the RegAddress in the Read section. Click the
Read Registers
button. Verify x8000 was written to this address. Close this window.
Setup Procedure