Basic Test Procedure
14
SLAU455B – August 2012 – Revised May 2016
Copyright © 2012–2016, Texas Instruments Incorporated
ADS58H4x (ADS58H40/ADS58H43) EVM
Figure 12. ADS58H4x 11-bit SNRBoost Mode, Single Tone Test Result: Fs = 245.76 MSPS, Fin = 170 MHz
Single tone FFT test (ADS58H4x 14-bit High Resolution Burst Mode)
1. Select the ADS58H40-14b device.
2. After selecting Single Tone FFT Test under
Test Selection
, enter the sample rate, input frequency, and
the number of samples.
3. Select channel A, B, C, or D depending on which channel is connected to the signal generator.
4. On the High Speed Data Converter Pro menu, select
Data Capture Options
→
Trigger Option
. Check
both
Trigger Mode Enable
and
Software Trigger Enable
.