background image

www.ti.com

Basic Test Procedure

Figure 4. TSW1200 GUI

2.5

ADS42xx Test Procedure

Switch on 5V power supply on ADC EVM. Verify source current is about 330 mA ± 30 mA with clock
and TSW1200 connected at EVM default (i.e., connected at 250MHz).

Connect clock signal at J19 through an appropriate BPF.

Adjust the signal generator amplitude output to 0.6Vrms to provide approximately 0.8 Vpp at J19
through a BPF with about 5 dB attenuation plus cable losses.

Use a high quality, low phase noise generator for this input to ensure proper device evaluation.

Connect input signal through an appropriate BPF at either J6 or J3 (Channel A or B, i.e., 170MHz).

Adjust frequency of generator to match the coherent frequency displayed in the TSW1200 GUI.

Select the proper Display Channel on the TSW1200 GUI software that corresponds to the input
connection.

Adjust the signal generator amplitude output to achieve –1 dBFS on FFT plot.

Use a high quality, low phase noise generator for this input to ensure proper device evaluation.

Initiate a capture by pressing the Capture button on the TSW1200 GUI.

7

SLAU333 – March 2011

ADS42xx EVM

Submit Documentation Feedback

© 2011, Texas Instruments Incorporated

Содержание ADS42 EVM Series

Страница 1: ...ults 8 3 Optional Configurations 8 3 1 THS4509 Input Op Amp Configuration 8 3 2 On Board CDC72010 Clock 9 3 3 Parallel CMOS Output 9 List of Figures 1 ADS58C28_ADS42xx_GUI Front Panel Top Level 3 2 AD...

Страница 2: ...as two register tabs Top Level and Advanced The Top Level tab provides an interface to the most used registers The Advanced tab includes less used registers and provides an option to manually input ad...

Страница 3: ...GUI to bring the advanced panel to the front On this panel user can write address and data of any register into the address data boxes as shown in Figure 2 45 is address and 4 is data Press Enter on t...

Страница 4: ...y updated o Select Load Regs button o Double click on the desired register file o Click on Send All to ensure all of the values are loaded properly 1 2 4 Miscellaneous Settings Device Selection Select...

Страница 5: ...Short 1 2 DC supply JP17 Short 2 3 DC supply LDO JP19 Short 2 3 DC supply LDO JP28 Short 2 3 DC supply LDO JP29 Short 2 3 DC supply LDO JP26 Open DC supply for ext buffer JP27 Open DC supply for ext...

Страница 6: ...the TSW1200 set up and operation This document assumes the TSW1200 software is installed and functioning properly The front panel of the TSW1200 is shown in Figure 4 The following configuration needs...

Страница 7: ...erator for this input to ensure proper device evaluation Connect input signal through an appropriate BPF at either J6 or J3 Channel A or B i e 170MHz Adjust frequency of generator to match the coheren...

Страница 8: ...input are shown in Table 3 Table 3 Jumper Setting for Transformer Coupled or OPA Driven Input Jumpers or 0 Transformer Coupled Input default OPA Driven Input R119 Install Do not install R123 Install D...

Страница 9: ...ault configuration see CDCE72010 data sheet SCAS858 for details about device default configuration With a 10MHz primary reference at J19 and a 983 04 MHz VCXO on board the CDC outputs a LVCMOS clock a...

Страница 10: ...ct This notice contains important safety information about temperatures and voltages For additional information on TI s environmental and or safety programs please contact the TI application engineer...

Страница 11: ...orized for use in safety critical applications such as life support where a failure of the TI product would reasonably be expected to cause severe personal injury or death unless officers of the parti...

Отзывы: