Description of the signal inputs and outputs
39
Maintenance message
If the LED above the
Alarm
key lights up yellow, maintenance carried
out by a service technician is required. The interval operation is con-
tinued. The respective message is recorded on the SD card. The
alarm relay does not switch.
The service technician determines which maintenance work has to be
performed by using a computer at the USB interface and then per-
forms the maintenance work.
When this has been completed, he or she acknowledges the mainte-
nance message by pressing the
Alarm
key for 3 seconds. Now, he or
she presses the
Manual
key and the device returns to the interval op-
eration.
Maintenance message
Description
Reaction of the
device
Measures
13 Service ex-
ceeded
The maintenance inter-
val set by the user has
been exceeded. The de-
vice must undergo
maintenance.
Interval opera-
tion is continued
After you have per-
formed the maintenance,
acknowledge the mainte-
nance message. A new
maintenance interval is
started.
25 Change pump
head 1
The operating time of
pump head 1 has been
exceeded. It must be re-
placed.
Interval opera-
tion is continued
After you have replaced
the pump head,
acknowledge the mainte-
nance message. As a re-
sult, the counter for the
pump run time is re-
started.
26 Change pump
head 2
The operating time of
pump head 2 has been
exceeded. It must be re-
placed.
Interval opera-
tion is continued
After you have replaced
the pump head,
acknowledge the mainte-
nance message. As a re-
sult, the counter for the
pump run time is re-
started.
Alarm and maintenance at the same time
If an alarm and maintenance work are both pending at the same time,
the
Alarm
key lights up red (for alarm) and the yellow LED lights up
above the key (for maintenance).
NOTE
lights up
pause
press 3
seconds
Acknowledge the
maintenance: