The Service Monitor program
28
Selecting language
“German”, “English”, “French” or “Dutch” can be selected as a menu
language under the “Language” tab.
Adjusting, saving an loading parameters
“Device selection” field
Select the connected device, e.g. Testomat
®
Modul CL.
Select device before loading the version
If you change the device selection after loading the version, the serial
interface is closed and the USB communication is terminated.
“USB” field
Select the port number used by the computer. It is assigned during
the USB driver’s installation.
“Date time” field
The system time is adopted from the connected computer with the
“Set time” button.
Automatic switching between summer and winter time can be acti-
vated with the “Summer/winter time automatically” check mark.
Date and time in messages
The date and time must be set to provide a time stamp in the history
of the error messages, maintenance messages and measured val-
ues. If the time is not set, the timestamp records
01.01.2011; 12:00
.
“Version” field
The version number and the creation date of the software used (boot-
loader and firmware) are read from the device and displayed using
the “Read” button.
“Settings” field
Set the operating mode, flush time, total operating hours and the time
to change the pump heads.
Operating mode (measurement phase mode/continuous mode)
You can choose between measurement phase mode and continuous
mode.
To set the measurement phase mode, adjust the “Measurement
phase” slide controller accordingly.
If you want to activate continuous mode, place the check mark next
to “Continuous mode”.
NOTE
NOTE