testo 6631 - Commissioning
12
Parameterizing/adjusting/analyzing the
instrument
The analog outputs can be scaled in the following ranges, beyond the
measurement range of the measurement transmitter: -50…100°C/
-58…212°F, -50…150%RH.
Scaling beyond the measurement range allows the adaptation of the
scaling limits to prescribed values from a PLC. The measurement range
itself cannot be extended this way.
The instrument is parameterized, adjusted and analyzed using the P2A
Software, see “Part 2: testo P2A Software”.
Alternatively, the 1-point adjustment can be carried out using the adapter
(accessory, 0554 6022) and a testo 400/650 with reference humidity probe.
1-point adjustment with adapter
1
Connect testo 400/650, with
reference humidity probe
(0636 9741) connected, to
the interface of the testo
6631 via the adapter
(connected to probe socket
1 of the testo 400/650).
2
Expose the testo 6631 and
the reference humidity probe
to the same reference
conditions (e.g. in the
humidity generator) and
allow climatic conditions to
equalize.
3
Switch on the testo 400/650. The values of the testo 6631 and reference
humidity probe are displayed on the left and right-hand sides of the two-
part display of the testo 400/650 respectively.
4
Adjust the humidity and temperature values on the reference humidity
probe using the
Probe
>
Probe Adjustment
menu item on the testo 400/650.
5
Disconnect the adapter from the interface of the testo 6631.
Содержание 6631
Страница 1: ...Instruction manual en testo 6631 Bio research transmitter P2A Software for testo 6631 ...
Страница 2: ...2 ...
Страница 8: ...testo 6631 Specifications 8 Dimensions ...
Страница 24: ...0970 6631 en 02 V01 00 V01 10 2 en ...