testo 6631 - Commissioning

10

Commissioning

Connecting instrument

²

Replacing vventilator: 

1.

Connect plug-in cable.

2

. Place ventilator drawer assembly

into housing.

3

. Tighten screw on ventilator drawer

assembly.

²

Install iinstrument iin aa ssuitable pposition:

1

Connect the supply line to the socket plug on the instrument.

Содержание 6631

Страница 1: ...Instruction manual en testo 6631 Bio research transmitter P2A Software for testo 6631 ...

Страница 2: ...2 ...

Страница 3: ...necting instrument 10 Electrical connection 11 Parameterizing adjusting analyzing the instrument 12 Maintaining the product 13 Tips and assistance 14 Part 2 testo P2A Software Specifications 15 First steps 16 Installing the software driver 16 Starting the software 16 Product description 17 Using the product 18 Tips and assistance 23 de fr es it pt sv nl ...

Страница 4: ...physical injury or damage to equipment if the precautionary measures indicated are not taken Important Avoiding personal injury damage to equipment i Never store the product together with solvents and do not use any desiccants i Only operate the product properly for its intended purpose and within the parameters specified in the technical data Do not use force i Only carry out the maintenance and ...

Страница 5: ...rch transmitter Long term stability and reliability thanks to the precise Testo humidity sensor Integrated ventilator allows targeted flow impact onto the sensor making the recording of mean climate conditions within greenhouse cells Time savings in commissioning and maintenance due to Parameterization adjustment and analysis software P2A Easy and fast ventilator replacement thanks to ventilator d...

Страница 6: ...Language versions of the instruction manual German English K02 Language versions of the instruction manual French English K03 Language versions of the instruction manual Spanish English K04 Language versions of the instruction manual Italian English K05 Language versions of the instruction manual Dutch English K06 Language versions of the instruction manual Japanese English K07 Language versions o...

Страница 7: ...r 4 wire Meas rate 1 s Voltage supply 24 V AC DC 10 AC Ambient conditions Operating temperature 10 60 C 14 140 F Storage temperature 20 70 C 4 158 F Housing Material ABS UV proof Colour white Dimensions 105 x 139 x 225 mm Weight 800 g Protection class IP66 Weight 1000 g Protection class transmitter IP 65 Protection class housing IP 33 Display optional 2 line LCD Directives standards and tests EC D...

Страница 8: ...testo 6631 Specifications 8 Dimensions ...

Страница 9: ...t a glance Suspension ring Voltage supply signal output Ventilator Ventilator drawer assembly rear Service opening for humidity sensor rear Mini DIN connection for P2A software and 1 point adjustment Two line display Ventilation shaft de fr es it pt sv nl ...

Страница 10: ...ument Replacing v ventilator 1 Connect plug in cable 2 Place ventilator drawer assembly into housing 3 Tighten screw on ventilator drawer assembly Install iinstrument iin a a s suitable p position 1 Connect the supply line to the socket plug on the instrument ...

Страница 11: ...AC ventilator 4 Not in use 5 Top view of plug socket DC Supply transmitter AC Supply ventilator 4 wire technology Plug manufacturer Firma Tuchel Amphenol Cable socket not included in delivery Type 01630D00610010 Pin socket fitted in instrument ex works Type Eco mate instrument plug 1 RH 4 20mA 2 C F 4 20mA 3 AC ventilator and transmitter 4 Top view of plug socket AC Supply transmitter and ventilat...

Страница 12: ...to 400 650 with reference humidity probe 1 point adjustment with adapter 1 Connect testo 400 650 with reference humidity probe 0636 9741 connected to the interface of the testo 6631 via the adapter connected to probe socket 1 of the testo 400 650 2 Expose the testo 6631 and the reference humidity probe to the same reference conditions e g in the humidity generator and allow climatic conditions to ...

Страница 13: ...olvents Cleaning t the s sensor During cleaning avoid touching the sensor at all costs 1 Remove service opening 2 Unscrew sensor protection cap 3 Carefully rinse the mirrored surface with isopropyl alcohol and or distilled water 4 Allow the sensor to dry completely 5 Unscrew sensor protection cover 6 Close service opening Exchanging s sensor 1 Loosen screw on the ventilator drawer assembly 2 Remov...

Страница 14: ...ter cable for USB to mini DIN 0554 6020 Adapter for 1 point adjustment with testo 400 650 0554 6022 Process display testo 54 2 AC two relay outputs to 250 VAC 300 VAC 3 A mains power supply 90 260 VAC 5400 7553 Process display testo 54 7 AC two relay outputs to 250 VAC 300 VAC 3 A mains power supply 90 260 VAC with RS485 output for Online Monitoring and with toatlizer display 5400 7555 Humidity ad...

Страница 15: ...yzing software for Testo transmitters It is not supplied with the testo 6631 System requirements Operating system Windows 2000 SP4 Windows XP or Windows Vista Computer Pentium processor of at least 400 MHz or equivalent 128 MB RAM Graphics resolution of at least 1 024 x 768 Unused hard drive capacity of at least 15 MB CD ROM drive USB interface de fr es it pt sv nl ...

Страница 16: ...a My Computer or Windows Explorer 2 Follow the instructions of the installation program Installing t the U USB d drivers The USB driver CD is supplied with the P2A Software Before installing the USB drivers please read the separate documentation enclosed with the USB driver CD The installation of the USB driver is the prerequisite for the faultless use of the P2A software P2A S Software u update 1...

Страница 17: ...Parameter files contain only parameter data These can be copied into another instrument or paramerter file of the same instrument type so that several instruments have the same parameter settings The name is freely selectable however it is recommended that the instrument context is retained Parameter files are always marked in red the parameter values contained in them are only passed on to the in...

Страница 18: ...connection w with t the d device Several instruments can be connected to the PC and administered via the P2A software but only one connection can ever be active at any one time Non wired instruments can also be connected to the P2A Software for parameterization adjustment The supply to the instruments is then effected via the USB interface analog outputs not functional 1 Connect the USB mini DIN a...

Страница 19: ...points of the scaling can be selected up to the stored scale minimum and maximum In order to adapt the analog output to the customer s system it is possible to scale beyond the end of the measuring range Unit Selection of the physical unit When the unit is changed standard values are set at Scale minimum and maximum Attention Changes to the physical unit cause the relay limit values to be set to t...

Страница 20: ... files stored in other directories can also be opened Only parameter data stored in the standard file can be loaded into an instrument 1 In the menu bar click on File Open 2 Select the storage location and click on the requisite parameter file 3 Click on Open to confirm entries Copying t the p parameter d data The parameter data for an instrument parameter file can be transferred to another instru...

Страница 21: ...on period to lapse 4 Enter reference value and perform adjustment by clicking on Carry out 1 point adjustment i To reset an offset value click on Set Offset to zero 5 Click on OK to confirm entries Carrying o out a a 2 2 p point a adjustment Before the 2 point adjustment it must be ensured that the ventilator of the testo 6631 is operating and that the air input opening is not blocked 1 Click on A...

Страница 22: ...tton and follow the instructions issued by the P2A Software 5 Click on OK to confirm entries Viewing a a t transmitter h history The current history data are always displayed as stored in the instrument file A distinction is made between parameterization and adjustment histories Dates and times refer to the PC time when the P2A Software was being used History data are only stored in the instrument...

Страница 23: ...ected 1 In the menu bar click on File Add Folder 2 Give the new folder a name Tips and assistance Questions and answers Question Possible causes solutions Connection to instrument cannot be established Check connecting cable and plug contacts USB driver not incorrectly installed Re install Adjustment is to be reversed Carry out factory reset Click on Test analyze transmitter Click on Carry out fac...

Страница 24: ...0970 6631 en 02 V01 00 V01 10 2 en ...