7 Maintaining the product
93
8.
Place the motor of the condensate pump into the bracket.
9.
Fasten the motor on the condensate pump (short clockwise
turn).
10.
Plug the inlet hose (length 25mm) and outlet hose (length
31mm) of the pump onto the connecting sockets of the meas.
box.
11.
Push the pump onto the motor shaft until the clip locks engage.
Make sure that the hoses are not pinched or constrained.
12.
Attach cover.
7.12.
Replacing the filtration non-woven in the gas
cooler
The filtration non-woven is included in the filter set 0554
3381
✓
The meas. box must be switched off and isolated from the
mains supply.
1.
Unlock the condensate trap and pull it vertically off the
measuring box.
2.
Pull off the hose.
3.
Open the cover of the filtration non-woven in anti-clockwise
direction.