5 First steps
44
> press
[ ]
.
- The rinse phase starts.
- The flue gas analyser switches off. It is normal for the fan of the
meas. box to run on for a while.
5.3.
Folders / Locations
(only available via
Meas. Box
tab)
All readings can be saved under the currently active location.
Readings not yet saved are lost when the measuring instrument is
switched off.
Folders and locations can be created, edited, copied and activated.
Folders and locations (incl. protocols) can be deleted.
Calling up the function:
>
[ ]
→
Folders
→
[OK]
.
Adapting the display
:
> Toggle between overview (display of number of locations per
folder) and detailed view (display of all locations per folder):
[Overview]
or
[Details]
.
Activating a location:
>
Select the location →
[OK]
.
- The location is activated and the menu
Measurement Type
is
opened.
Creating a new location
:
A location is always created in a folder.
1.
Select the folder in which the location is to be created.
2.
[Options]
→
New location
→
[OK]
.
3.
Enter values or make settings.
The following inputs/settings are possible:
Parameter
Description
Location
Enter name
Application
Select application
Fuel
Select fuel
Profile
Enter diameter, length, width, height and area.
For correct measurement of the volume flow you
must set the profile and area. A volumetric flow
rate is calculated from the geometries entered
here together with the measured velocity.