Brief instructions
en
testo 330 -1
, -2
Flue gas analyser
These brief instructions provide an overview over the most important functions of the
product. Before using the product you should absolutely read and follow the operating
instructions! This document describes the products testo 330-1 LL and testo 330-2 LL
with the instrument setting Country version | Great Britain.
Device connections
➀
Sensor socket: Connect the sensor before
switching the measuring instrument on or, after
changing a sensor, switch the instrument off and
on again.
➁
Flue gas socket: It is possible to change a probe /
sensor even while the measuring instrument is
switched on.
➂
Mains unit socket
➃
Pressure socket
Replacing the probe module
1
Press the key on top of the probe handle and
remove the probe module.
2
Plug in the new probe module and engage it in
place.
Emptying the condensate trap
The condensate consists of a weak mix of acids. Avoid contact with the skin. Make
sure that the condensate does not run over the housing.
Do not drain the condensate trap
while the flue gas pump is running!
1 Hold the instrument with the condensate outlet
pointing up.
2 Open the condensate outlet on the condensate
trap: Pull out approx. 5mm against the stop (
➀
).
➀
➁
➃
➂
➀