Checking the instrument
18
Maintaining the product
Condensate trap
The fill level of the condensate trap can be read from the markings on the
condensate trap.
E
Em
mppttyyiinngg tthhee ccoonnddeennssaattee ttrraapp
Flue Gas condensate consists of a weak mix of acids. Avoid contact with the
skin. Make sure that the condensate does not run over the housing.
Caution! Failure to empty the condensate trap when the fill level is
reached results in condensate entering the sensors and pump. This
voids the warranty and results in costly repairs and replacements.
Do not empty the condensate trap while the flue gas pump is in
operation!
11
Keep measuring instrument in an upright
position (condensate outlet pointing upwards).
22
Open the condensate outlet on the
condensate trap: pull out approx. 7 mm to
the stop.
33
Let the condensate run out into a sink.
44
Mop up any remaining drops on the
condensate outlet using a cloth.
55
Close the condensate outlet.
The condensate outlet must be completely
closed, otherwise measuring errors could occur if external air gets in. It is
advisable to empty water trap at regular intervals, (preferably daily).
P
Prriinnttiinngg//ddeelleettiinngg//ddiissppllaayyiinngg m
meeaassuurreem
meenntt ddaattaa ((332277--22 oonnllyy))::
There are 20 memory locations (P
Poossiittiioonn 11 to P
Poossiittiioonn 2200) to which one
measurement data record can be saved per location. Memory locations
that have already been assigned are indicated by the display of the
date/time of saving.
11
Press .
22
Select M
Meem
moorryy using / .
- The memory capacity and available memory locations are displayed.
To print the memory: press the P
Prriinntt function key.
To delete the whole memory: press the D
Deell function key and confirm entry
with the Y
Yeess function key.
33
Press O
OK
K.
44
Select memory location using / .
To display the measurement data record: press the V
Vaalluuee function key.
To print the measurement data record: press the P
Prriinntt function key.
To delete the measurement data record: press the D
Deell function.
TTrraannssffeerr ddaattaa ttoo aa P
Poocckkeett P
PC
C ((332277--22 oonnllyy))::
Data can be transferred to a Pocket PC via infrared or Bluetooth.
You must also refer to the documentation that comes with the software.
Checking the instrument
P
Peerrffoorrm
miinngg aann iinnssttrruum
meenntt ddiiaaggnnoossiiss::
11
Press .
22
Select D
Diiaaggnnoossiiss using / and confirm entry with the O
OK
K function key.
33
Select the required function using / and confirm entry with the O
OK
K
function key.
FFuunnccttiioonnss
11.. IInnffoo: displays instrument information: serial number, instrument temperature, operating
hours, qA version, last service
22.. E
Errrroorr: displays list of errors.
33.. R
Reecchh.. bbaatttt: displays the battery capacity.
44.. S
Seennss.. D
Diiaagg (327-2 only): performs sensor diagnosis.
K
Keeyy ffuunnccttiioonnss ffoorr S
Seennss.. D
Diiaagg ffuunnccttiioonn ((332277--22 oonnllyy))
·
: select sensor.
· R
Reeaadd: performs sensor diagnosis and displays results of diagnosis.