Multi Station
Vicat Test Apparatus
42
5.5 Cleaning the filter
The procedure is the same for all apparatuses (4, 6, 8 and 12 stations).
Figure 46: cleaning the filter
Note
The filter sponges must be wet before inserting, so that no air drawn from the
sponges into the system.
If there is no water flow into the filter or the pump does not promote water, please
turn off temperature controller switch, open both ball valves and drain some water to
carry out any air bubbles from the system. Then close the ball valves and turn on
temperature controller switch again. (Repeat above procedure if necessary.)
Fill the basin with water before inserting the filter
cartridge. The screen must be dry to prevent the for-
mation of air bubbles under the screen!
3
5
Clean screen and
dry it.
4
Fix knurled
thumb screws
und
vent the air.
6
1
Loosen the knurled
thumb screws.
2
Remove the fil-
ter cartridge.
Remove and
wash sponges.
Order of Insertion:
1. filter-cartridge
2. fine-sponge
3. coarse-sponge