Laboratory mortar mixers
Model 1.0204.LV
19
7.6 Correct distance
If the distance between stirrer and mixing bowl is not correct,
remove stirrer and mixing bowl.
First, unscrew the cover.
Slightly loosen the two clamping screws.
Now loosen the lock nut of the adjusting screw at the lower end of the mixing bowl
holder, and then set the distance with the help of the adjusting screw from the mixing
bowl mounting and retighten the clamping screws.
Check the distance with the spacing template.
By turning the adjusting screw in a clockwise direction, the distance is reduced.
By turning the adjusting screw in a counterclockwise direction, the distance is in-
creased.
Insert bowl and stirrer and check the distance, repeat the process until the dis-
tance is correct.
If the distance is set correctly, the lock nut, the adjusting screw and both of the
clamping screws on the holder for the mixing bowl must be tightened
firmly
. The
contact protection must be reassembled, the mixing bowl and the stirrer must
then be inserted.
Figure 6 Height adjustment
Remove these screws
Remove
the cover
Clamping screw
Clamping screw
Lock nut
Adjusting screw