Product Specification
Product
TSR4 WLAN Wireless Radon Probe
Type symbol
042 127 210 002; 042 127 285 000
Average measurement sensitivity
0,25 count/hour/Bq.m-3
(method RaA+RaC; 15°C ÷ 30°C; rel. hum. 20% ÷ 40%)
Measuring range
MDA – 100 000 Bq/m
3
;
MDA = 100 Bq/m
3
per 1 hour or 20 Bq/m
3
per 24 hours
Measurement uncertainty
< 13% at 300 Bq/m
3
per 1 hour;
< 3% at 300 Bq/m
3
per 24 hours
Measuring chamber capacity
0,176 dm3
Response rate
< 30 minutes (RaA); < 3 hours (RaA + RaC)
Radon records
calculated from RaA (quicker, less sensitive)
calculated form RaA + RaC (slower, more sensitive)
Measuring relative humidity range
0 – 100 %
Measuring temperature range
-40 to + 125 °C
Radio interface
868MHz
Max number of measuring network
elements
16
Probe to terminal unit RF range
depends on number of walls and building material, up to 600 m in
open space
Possibility of using repeater
yes
Records saving interval
1- 255 minutes, default 1 hour
Results memory capacity
5000 records (208 days of 1 hours records), spectra 300 records
Powering
internal rechargeble accumulator; charging via USB
Accu life after full charging
>1 year (also depens on operating conditions)
Current radon concentration
short-term (0,5 hour running average from RaA)
long-term (24 hours running average from RaA + RaC)
Dimension
Ø 80 x 175 mm
Waterproof
IP67 (only for electronics)
Operating conditions
Temperature: -10 ° C to +40 ° C
Recommended relative humidity: 10% - 75%
Maximum working relative humidity: 0% - 99%
* Increased humidity reduces the life of a charged battery.
* There must be no condensation of water in the chamber -
erroneous results
Detector life
50-100 million pulses;
that means at an average concentration of 1000 Bq / m3 -> 12
years;
10,000 Bq / m3 -> 1 year
5
Operating Instructions
Switching on and off:
The probe measures radon concentration autonomously and communicates in wireless network only if
the switch is in position „on“. The switching on is signalized by LED diode „STAT“ according chart below.
If the switch is in position „off“ the probe doesn´t measure radon concentration and doesn´t
communicate in wireless network. In switching off mode the probe only keeps running real time for correct
date and time of records in case of switch it on again. By switching off the probe doesn´t lose previous
records of measurement. The switching on is signalized by LED diode „STAT“ according chart below.
It is possible to download data from probe over USB in both position of switch.