background image

 

6/8 

4

 

Product Specification 

 

Product 

TSR4 WLAN Wireless Radon Probe 

Type symbol 

042 127 210 002; 042 127 285 000 

Average measurement sensitivity 

0,25 count/hour/Bq.m-3 
(method RaA+RaC; 15°C ÷ 30°C; rel. hum. 20% ÷ 40%) 

Measuring range 

MDA – 100 000 Bq/m

3

;  

MDA = 100 Bq/m

3

 per 1 hour or 20 Bq/m

3

 per 24 hours 

Measurement uncertainty 

< 13% at 300 Bq/m

3

 per 1 hour; 

< 3% at 300 Bq/m

3

 per 24 hours 

Measuring chamber capacity 

0,176 dm3 

Response rate 

< 30 minutes (RaA); < 3 hours (RaA + RaC) 

Radon records 

calculated from RaA (quicker, less sensitive) 

 

calculated form RaA + RaC (slower, more sensitive) 

Measuring relative humidity range 

0 – 100 % 

Measuring temperature range 

-40 to + 125 °C  

Radio interface 

868MHz  

Max number of measuring network 
elements 

16  

Probe to terminal unit RF range 

depends on number of walls and building material, up to 600 m in 
open space 

Possibility of using repeater  

yes 

Records saving interval 

1- 255 minutes, default 1 hour 

Results memory capacity  

5000 records (208 days of 1 hours records), spectra 300 records 

Powering 

internal rechargeble accumulator; charging via USB 

Accu life after full charging 

>1 year (also depens on operating conditions) 

Current radon concentration  

short-term (0,5 hour running average from RaA) 

 

long-term (24 hours running average from RaA + RaC)

 

Dimension 

Ø 80 x 175 mm 

Waterproof 

IP67 (only for electronics) 

Operating conditions 

Temperature: -10 ° C to +40 ° C  
Recommended relative humidity: 10% - 75%  
Maximum working relative humidity: 0% - 99%  
* Increased humidity reduces the life of a charged battery. 
* There must be no condensation of water in the chamber - 
erroneous results 

Detector life 

50-100 million pulses; 
that means at an average concentration of 1000 Bq / m3 -> 12 
years;  
10,000 Bq / m3 -> 1 year 

 
 

5

 

Operating Instructions 

 
Switching on and off: 

   The probe measures radon concentration autonomously and communicates in wireless network only if 

the switch is in position „on“. The switching on is signalized by LED diode „STAT“ according  chart below.  

 If  the  switch  is  in  position  „off“  the  probe  doesn´t  measure  radon  concentration  and  doesn´t 

communicate in wireless network.  In switching off mode the probe only keeps running real time for correct 
date  and  time  of  records  in  case  of  switch  it  on  again.  By  switching  off  the  probe  doesn´t  lose  previous 
records of measurement. The switching on is signalized by LED diode „STAT“ according  chart below.   

It is possible to download data from probe over USB in both position of switch.   

 

Содержание TSR4A

Страница 1: ...Specifications Operation Manual v 1 2022 Table of Contents 1 Introduction 2 2 Description and Utilization 2 3 Scope of Delivery 5 4 Product Specification 6 5 Operating Instructions 6 6 Basic Maintenan...

Страница 2: ...used product If you experience any problems with your product please contact us at TESLA Rubeska 215 1 190 00 Prague 9 Vysocany www tesla cz 1 Introduction This document describes technical specifica...

Страница 3: ...sla cz TSR4A Radon Probe can be operated by these ways A Standalone probe Thanks to its independent battery power portable radon measuring probe supports flexible placing options within monitored stru...

Страница 4: ...centration values to central unit wirelessly Central unit analyzes this information and on the basis of measured set concentration level value it sends command to actuator wireless actuator or actuato...

Страница 5: ...5 8 3 Scope of Delivery TSR4A WLAN Wireless Radon Probe Power adapter 230VAC 5VDC USB cable A B Antenna Probe holder Operation Manual...

Страница 6: ...on operating conditions Current radon concentration short term 0 5 hour running average from RaA long term 24 hours running average from RaA RaC Dimension 80 x 175 mm Waterproof IP67 only for electron...

Страница 7: ...way of radon probe If the probe is used for autonomous measurement and data downloading via USB interface then the setting and configuration are realized by usbRADONview application UsbRADONview appl...

Страница 8: ...recalibration is free from the manufacturer 7 EC Declaration of Conformity EC Declaration of Conformity will be delivered by Tesla producer on request If interested please use contacts on the web www...