Check the voltage level under the following conditions
Air-discharge
Single discharge
Polarity: positive and negative
Voltage settings: 2, 4, 8 and 15 kV
Permissible deviation: < ±5% of the set value
Check the discharge current pulse shape.
Check to be made under the following conditions
Contact-discharge
Single discharge
Polarity: positive and negative
Voltage settings: 2, 4, 6 and 8 kV
Compare the results obtained against the reference figures quoted
in IEC/EN 61000-4-2, Ed. 1.2:2001
Remark
Teseq offers an accredited service for this kind of work.
100%
90%
I at 30 ns
I at 60 ns
10%
t
r
= 0.7 to 1.0 ns
30 ns
60 ns