44
NSG 435 ESD simulator
8.1 Servicing
Care
The housing may be cleaned with a moist cloth with possibly just a trace of
detergent liquid.
Industrial spirit is also a suitable cleaning agent.
Other solvents are not permitted.
Fuses
The instrument contains no fuses that are accessible to the user.
8.2 Calibration
Trimming procedures in the NSG 435 are carried out digitally and automatically.
The instrument contains no elements that are foreseen for adjustment by the
user. A component defect must be suspected if the calibration measurements
differ from the published technical data and the instrument is to be returned
to an authorized Teseq service centre.
Measurements can only be undertaken by trained specialists. A prerequisite is the
availability of the necessary measurement equipment as listed in section 7.
Charge voltage check
Equipment:
EHT voltmeter with 20 kV voltage range
Internal resistance > 20 GΩ
Measurement accuracy <1%
8 MAINTENANCE