NSG 435 ESD simulator
Use F5 to return to the original menu level. (This return also occurs automatically
after about 10 s). The «Automatic polarity change» function is identified by «ALT.
POL.» over F4 and +/- over F3.
The instrument switches over from positive to negative polarity once half of the
preset number of pulses has been released (the sign on the display changes).
This automatic function works in both the single pulse mode as well as with
repetitive discharges.
Reinitialize the operation in single pulse mode after each cycle. Reset the
counter and select the counter menu again (press F5 three times).