27
Note: One battery charge will last for several days of normal test operation.
The useful working time depends, of course, to a large extent on the operating
conditions.
The following figures have been obtained by way of reference:
Battery freshly charged
Air-discharge at 15 kV
More than 10000 discharges until «LOW BAT» blinks
A further 3000+ discharges until «LOW BAT» remains constant
5.3 Battery monitor
The battery charge state is monitored continuously. An insufficiently charged
or empty battery is indicated on the display.
«LOW BAT» blinks:
Change or charge the battery soon. The in-
struments functions and the pulse data are still
assured.
«LOW BAT» steady:
Battery empty. The instrument’s functions are
inhibited since the pulse data can no longer be
assured.