21
3.3.6 Remote triggering
A remote triggering unit can be connected to operate the NSG 435 inside
a Faraday cage with external pulse triggering or for test pulse triggering in
synchronism with other conditions. The optical input for remote triggering is
insensitive to external light sources.
The normal remote trigger unit consists of a «triggerbox» and 5 m (197 “) of
optical cable. The remote trigger works in parallel with the «trigger-button» on
the NSG 435. Pulse triggering, or the on/off switching in the case of repetitive
discharges, can be effected by a push-button or an electrical signal applied to
the triggerbox.
The electrical signal at the BNC connector must fulfill the following condi
-
tions:
On: V = 2.4...10 V, I >2 mA, t >10 ms
Off: V <0.8 V
Repetition rate: <5 Hz
The remote trigger is powered by a conventional 9 volt battery. The current
consumption is so low that a battery switch has been dispensed with. If NSG
435 provided with two optical connectors connect the optical cable to the blue
terminal.
It is recommended to remove the battery if the unit is not going to be used for
a long time.