20
NSG 435 ESD simulator
The power supply unit must be connected to a mains outlet having a protective
earth.
The protective earth connection does not replace the earth cable for the opera-
tion of the NSG 435. To ensure safe and valid test operation the earth cable must
be correctly connected as the pulse return path in every case.
The mains power supply unit is constructed in conformity with the relevant
safety standards and carries the appropriate test symbol.
3.3.5 Discharge networks
The basic set contains a discharge network and test fingers that conforms to
IEC/EN 61000-4-2, Ed. 1.2:2001.
Alternative networks can be installed for testing in accordance with other
standards.
The discharge network and test fingers form a mutually matched combination.
They are labeled with a corresponding INA number. The specified pulse data
are only achieved while this combination is maintained.
Several combinations are given in the order list. The C and R values of the
discharge network can also be specified for other applications.
Networks conforming to other standards can be built upon request. The speci-
fications of the standard must be fully defined.
Exchanging the discharge network is described in section «Exchanging the R/C
network».