18
NSG 435 ESD simulator
3.3.2 Charging unit/battery-pack
The battery charger supplied in the set is designed for multi-country use and
is coming with various mains adaptors. Green LED shows fully charged battery
status.
Operation range 100 V to 240 V (50/60 Hz)
Charging of the battery takes about 1 hour. For further details please refer to
separate manual.
The battery has to be removed from the NSG 435 for charging via the jack con-
nector. One charge is sufficient for several days of normal test operation.
Battery life expectancy:
Ambient temperatures over 50°C can lead to degradation of the battery. If
treated carefully, more than 300 charge/discharge cycles can be expected
without any noticeable reduction in capacity.
The charger and battery-pack form a matched entity. The battery must not be
charged from any other unit and the charger is to be used exclusively for the
intended purpose.
Operating advice:
Avoid short-circuits. Keep the contacts clean
Use the equipment only in dry surroundings
Do not switch the mains off while charging otherwise the timer will be
restarted
Never leave the battery in a fully discharged state for a long time
Do not attempt to recharge a full battery
Recharge the battery about every 6 months even if the instrument is not
being used