Shield Box Operation without Fixture
27
4.5 Shield Box Operation without Fixture
In order to work Shield Box without fixture, the fixture mode must be set to 'Off Mode.'
Refer to
5.7.2.2 Probe Configuration Commands
If the fixture mode is set to 'Sensor Mode,' sensors will detect nothing and thus the lid will stay open as
if there is a problem..
Also, if it is set to 'Timer Mode,' the lid will take certain time to open and close, hence the operation is
delayed.