Instruction manual for MH & MH+M
95
15
MAX, MIN, Δ MODE
15.1 Introduction
This measurement mode is also called “continued display”. It can be defined as a mode that
enables scanning a surface in order to detect parallelism errors in regards to a reference
surface.
This mode is accessible by pressing the key
in
ST1
or
ST2
measurement
modes.
Menu
FX
from
ST1
Menu
FX
from
ST2
15.2 Fine adjustment
The fine adjustment device is used to precisely adjust a height. A manual gauge without
fine adjustment device can be retrofitted with such option using a special kit.
For questions please contact your local representative.
15.3 Measurement
principle (MH+M)
1. Once the measurement mode is activated, position the probe above the surface to be
measured.
2. Press the key
or
that corresponds to the desired measurement direction. The
probe will move in the chosen direction in order to establish contact with the workpiece
to be measured.
Содержание MICRO-HITE
Страница 10: ...Instruction manual for MH MH M 10 Fig Description of the constitutive elements of the TESA MICRO HITE M ...
Страница 83: ...Instruction manual for MH MH M 83 ...
Страница 99: ...Instruction manual for MH MH M 99 ...
Страница 152: ...Instruction manual for MH MH M 152 ...
Страница 163: ...Instruction manual for MH MH M 163 EXAMPLE TESA WORKPIECE ...