Door System Troubleshooting
The intent of centrifuge door and door lock system troubleshooting is to inform the technician about
where and when to measure values to help troubleshoot door position and door lock system
malfunctions and how to confirm the suspected error by testing.
Prerequisites
Check the following electronic prerequisites for normal operation:
•
24 V DC source from the power supply must be within the 19.2 V DC to 28.8 V DC range
(measure between J6 and J2 on the motor driver CCA, or check LED D14)
•
24 V DC switched from the motor driver CCA must be within the 19.2 V DC to 28.8 V DC range
(measure between J3 and J2 on the motor driver CCA, or check LED D1)
•
Safety CCA reset of 24 V shutdown (digital) indicates run or no reset (low)
•
Safety CCA must activate the 24 V DC switch (digital) on the motor driver CCA
•
24 V current sense line (DC voltage) from the motor driver CCA must not trigger the solenoid fault
line
•
Sensor LED power must be within range; 5.0 V DC maximum (switched on the safety CCA)
•
Lock solenoid power must be enabled
Figure 4-6: Door lock troubleshooting diagram
Door System Troubleshooting
Trima Accel
®
Automated Blood Collection System • Service Manual
4-17
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...