*versions 5.1.8 and 5.1.9
Exit Condition
When the valves are detected in the commanded position within 10 seconds.
Evacuate Bags Substate
This substate removes air from the product bags. The return pump runs until a predetermined volume
has been pumped (version specific). If the access pressure exceeds the alarm limit during this test, an
alarm is generated and the Trima Accel system reverts to the previous substate. If the test passes, the
Trima Accel system advances to the Lower Notification substate.
Table 3-8: Evacuate Bags Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC Valve
Centrifuge
(rpm)
0
0
0
0
90
Open
Open
Open
Return*
0
*versions 5.1.8 and 5.1.9
Exit Condition
When 140 mL is processed by the return pump.
Lower Notification (Cassette Position Message) Substate
This substate informs GUI of the cassette position (it should be down) to prepare for disposable tests. If
the cassette is down, the Trima Accel system advances to the Disposable Tests substate.
Table 3-9: Lower Notification Substate
AC Pump
(mL/min)
Inlet Pump
(mL/min)
Plasma
Pump
(mL/min)
Platelet
Pump
(mL/min)
Return
Pump
(mL/min)
Platelet
Valve
Plasma
Valve
RBC
Valve
Centrifuge
(rpm)
0
0
0
0
0
Open
Open
Open
0
Exit Condition
When the cassette down message is sent.
Disposable Tests State
The Disposable Tests state performs tests designed to determine whether the disposable tubing set is
properly loaded and functioning.
The disposable tests include the following activities:
•
Checks for the correct tubing set and the proper loading of the disposable
•
Checks the integrity of the set. The access pressure sensor checks both positive and negative pressure
of the disposable set to ensure its usability
•
Checks for AC, inlet, and return pump tubing occlusions
•
Checks for proper access pressure sensor function
Software Description
3-14
Trima Accel
®
Automated Blood Collection System • Service Manual
Содержание Trima Accel
Страница 1: ...Trima Accel Automated Blood Collection System Service Manual...
Страница 2: ...Trima Accel Automated Blood Collection System Service Manual Part No 777095 548 2015 09...
Страница 10: ...1 Operational Description Trima Accel Automated Blood Collection System Service Manual 1 1...
Страница 62: ...2 System Description Trima Accel Automated Blood Collection System Service Manual 2 1...
Страница 100: ...Linear Actuator System Trima Accel Automated Blood Collection System Service Manual 2 39...
Страница 133: ...System Description 2 72 Trima Accel Automated Blood Collection System Service Manual...
Страница 134: ...3 Software Description Trima Accel Automated Blood Collection System Service Manual 3 1...
Страница 261: ...Software Description 3 128 Trima Accel Automated Blood Collection System Service Manual...
Страница 262: ...4 Troubleshooting Trima Accel Automated Blood Collection System Service Manual 4 1...
Страница 288: ...5 Maintenance and Calibration Trima Accel Automated Blood Collection System Service Manual 5 1...
Страница 303: ...Maintenance and Calibration 5 16 Trima Accel Automated Blood Collection System Service Manual...
Страница 304: ...6 Specifications Trima Accel Automated Blood Collection System Service Manual 6 1...
Страница 333: ...Specifications 6 30 Trima Accel Automated Blood Collection System Service Manual...
Страница 339: ...Index 6 Trima Accel Automated Blood Collection System Service Manual...