7.1
Device
settings
These are the device-dependent settings, meaning they cannot be copied to other devices.
When using a TOMEs network, it is possible to alter them via TOMEs.
Following device settings are available:
Serial number
Information
Serial number of the device. This setting cannot be changed by the user.
Firmware v
Information
Version of the firmware. This setting cannot be changed by the user.
Service (don)
Information
Number of donations which have been performed since last servicing. This setting cannot
be changed by the user.
Password
0000
–
9999
Default: 9876
Allows to change the password for entering the settings.
Service (mon)
Information
Number of months since last servicing. This setting cannot be changed by the user.
Language
DAN / DEU / ENG / ESP / FRA / ITA / NED / NOR / POR / RUS / SUO / SVE / MAG /
POL / BRA / TUR
Default: ENG
Language of the firmware.
Time
00:00
–
23:59 (hh:mm)
Sets the time of the internal clock.
When T-RAC II is connected to TOMEs via the network,
the time is automatically taken from TOMEs. In that case, this setting cannot be changed.