Smart® ECM

 

Fan/Filter Unit  

with Integrated UPS Battery 

      

 

Quick Start Guide  

Doc. 1788-47

 

© 

Copyright 2021 Terra Universal Inc.   

All rights reserved 

 

 

 

Terra Universal, Inc.  •  TerraUniversal.com  •  800 S. Raymond Ave.  •  Fullerton, CA  92831  •  TEL: (714) 578-6000  •  FAX: (714) 578-6020  

1

 

IMPORTANT SAFETY INSTRUCTIONS 

 

READ AND SAVE THESE INSTRUCTIONS 

Proprietary Notice

 

This 

manual 

pertains 

to 

proprietary  devices  manufactured 

by  Terra  Universal,  Inc.  Neither 

this document nor any portion of it 

may  be  reproduced  in  any  way 

without  prior  written  permission 

from Terra Universal.

 

Safety Notice

 

A thorough familiarity with all operating guidelines is 

essential to safe operation of the product.  Failure to 

observe  safety  precautions  could  result  in  poor 

performance,  damage  to  the  system  or  other 

property,  or  serious  bodily  injury  or  death.  The 

following symbols are intended to call your attention 

to two levels of hazard involved in operation.

 

Terra  Universal  makes  no 

warranties applying to information 

contained  in  this  manual  or  its 
suitability  for  any  implied  or 

inferred purpose.  Terra Universal 

shall  not  be  held  liable  for  any 

errors this manual contains or for 

any  damages  that  result  from  its 

use. 

 

CAUTION

 

Cautions are used when failure to 
observe  instructions  could  result 
in 

significant 

damage 

to 

equipment.

 

 

WARNING

 

Warnings are used when failure to 
observe 

instructions 

or 

precautions  could  result  in  injury 
or death.

 

The information presented here is subject to change without notice.

 

 

FFU Specifications 

Size: 

2’x4’ 

Housing:  

White powder coated cold-rolled steel or 304 stainless steel 

HEPA Filter: 

99.97% efficient on removal of particles 0.3 microns and larger. 

Pre-Filter: 

20” x 20” x 1” – 30% efficient ASHRAE rated. 

Impeller: 

Forward-curved centrifugal type factory balanced. Entire 
motor/blower assembly is removable from top of housing for service. 

Face Grille: 

Perforated anodized aluminum or 304 stainless steel 

Velocity: 

90 FPM or higher 

Noise Level: 

40 dBA (measured at 30” from filter face) 

Weight: 

170 lbs. shipping weight 

Power:  

1/3 HP motor; 120V/230V, 50/60Hz, 4.128 Amps 

 

Specification Notes: 
•         All FFU data is based on a standalone unit using 1” prefilter and clean filters.  

•         Data will vary depending on filter media and configurations with other 

products/systems, such as ductwork or hoods.

 

 

1.0

 

Installation 

The Fan/Filter Unit is designed to fit on top of a Terra standard 2’ x 4’ clean room ceiling 

grid. It includes a threaded fixture at each corner to accommodate four eye-bolts (not 
provided) that can be used to assist in positioning the unit. 

 

WARNING: To reduce the risk of fire, electric shock, or injury to persons, 

observe the following: 

 

A)

 

Use this unit only in the manner intended by the manufacturer. If you have 

questions, contact the manufacturer. 

B)

 

Before servicing or cleaning unit, or replacing a filter, switch power OFF at the 

service panel and lock the service disconnecting means to prevent power from 
being accidentally switched on. When the service disconnecting means cannot be 

locked, securely fasten a prominent warning device, such as a tag, to the service 
panel. 

C)

 

When the removal/disconnection of either filter is required due to service or 
component replacement, the replacements are to be remounted as previously 

installed. 

D)

 

Installation work and electrical wiring must be done by qualified person(s) in 

accordance with all applicable codes and standards, including fire-rated 
construction. All metal components must be properly grounded. 

E)

 

When cutting or drilling into a wall or ceiling, do not damage electrical wiring and 
other hidden utilities. 

 

2.0

 

120V vs. 220V Configuration

 

 

CAUTION:

 

Double  check  your  FFU  is  configured  for  the  correct  voltage. 

Connecting  the  motor  to  the  wrong voltage can  permanently  damage  and 
break  the  motor.  The  FFU  is  pre-configured  for  120V.  It  can  be  easily 

reconfigured to 230V. 

 

3.0

 

Initial Start-Up 
WARNING: Keep unit disconnected from power supply during inspection. 

 

Prior to providing power to the unit, check that no damage has occurred during shipping. 

Conduct a visual check to make sure there are no visible dents or penetrations. If the 
unit is intact, remove the pre-filter and fan guard screen. Once the battery is plugged in, 

it will automatically start charging and take 1.5 – 2 days to fully charge. 

 

To  start  unit,  plug  the  UPS  into  a  live  AC  wall  outlet  and  then  press  and  hold  the 

POWER

 button for about one second. If there is no utility power, you can “cold-start” the 

UPS (turn it ON and supply power for a limited time from its batteries) by pressing and 
holding the

 POWER

 button for about two seconds (alarm will beep once after interval 

has passed). 

 

Do not unplug the UPS to test its batteries. This will remove safe electrical 
grounding  and  may  introduce  a  damaging  surge  into  the  network 

connections.

 

 

4.0

 

Operation 

The unit consists of a 1/3-HP electronically commutative motor driving a forward-curved 
centrifugal impeller. Air is drawn into the unit through a pleated pre-filter and in turn to 

an  insulated  chamber  where  it  is  then  directed  to  an  exit  HEPA  filter  with  a  built-in 
diffuser. 

The battery system provides approximately 1-2 hours runtime depending 

on FFU speed setting

. The external battery can be “hot-swapped” (safely removed or 

swapped without shutting down the UPS). 

 

The unit can also be equipped with a 3/8”-diameter challenge port, which can be used to 

take differential pressure measurements to monitor filter performance or insert aerosols 
for leak-test certification. Port is capped when not in use.  

 

120V Configuration 

1.

 

Black

 and 

Red

 wires are connected to the same terminal. 

2.

 

Yellow Jumper Cable is installed. 

 

220V Configuration 

1.

 

Black 

and

 Red

 wires connect to the upper terminal as shown. 

2.

 

Yellow Jumper Cable is removed. 