DPTR Rail Mount Differential Pressure Transmitter-Operation Manual Edition: A/3
1
DPTR Rail Mount Differential Pressure Transmitter
Applications & Features
Mount on standard 35mm DIN rail, small enough for
compact installation in equipments and instruments
Apply high accuracy MEMS sensor and digital technologies,
can measure differential pressure on equipment such as
isolators and clean benches for pharmaceutical production,
biological and medicine research, microelectronics,
hazardous chemicals operations, and general laboratories
Multiple ranges, engineering units and outputs
Good performance, accuracy 1.0% or 0.5%FS selectable
Function keys: zero reset, unit select, response time set,
etc
Specifications
Medium:
non-combustible, non-corrosive air, insensitive to
moisture, dust, condensation and oil
Temp.:
Working,-20~70°C; Medium, 0~60°C; Compensation, 0~50°C
Storage: -30~70°C
Working Pressure:
overload 10xFS, burst 15xFS
Performance:
Accuracy %FS
±1.0%
±0.5%
Thermal
effect %FS/°C(Zero/FS)
±0.05/0.08
±0.03/0.04
Stability %FS/Year
±0.5
±0.3
Response Time:
0.5/1/2/5s, can be set by keys
Process Connection:
5mm ID tubing
Output:
0~10V, 4~20mA (2 wires), RS485 selectable
Output Load:
≤500Ω (current), ≥2kΩ (voltage)
Power:
Current: 18.5~35VDC (R
L
=500Ω)8.5~35VDC (R
L
=0Ω)
Voltage: 16~28VAC/ 16~35VDC
Zero reset:
outside key to operate easily
Materials:
Fire-proof ABS+PC
Protection:
IP30
Weight:
110g
Approval:
CE
Models
Model
DPTR
Rail Mount DP Transmitter
Accuracy
0
±1%FS
1
±0.5%FS
Range
x
Range selection(See table)
Output
1 0~10V
2 4~20mA(2 wires)
8 RS485/Modbus
Measuring ranges
Note:
For zero center models, add “Z” at the end of the model. For example,
DPTR*1*Z, means the range is -30-0-30.Only 1~6 have this selection.
Connection
Different models have different electrical terminals. Please wire specific model according to the wiring diagram on the front
cover.
Setting
1. The zero-reset button is for resetting zero drift. Press and hold the button for 5 seconds to complete the operation.
2. The RS485 terminating resistor (120Ω) is selected by the toggle switch. When “J8” is ON, the resistor (120Ω) is applied.
3. Response Time select: switch the S2 to different positions for 0.5s,1s, 2s and 5s according to the diagram below.
Note: since the time is not fully synchronized, the actual response time is 1~2 times set time.
Code
Pa
Code
Pa
0
0
~
25
5
0
~
1000
1
0
~
60
6
0
~
2500
2
0
~
125
7
0
~
5000
3
0
~
250
8
0
~
10000
4
0
~
500
RS485
9600,8,N,1
Slave Add.= 1
3 4
A
1 2
POWER SUPPLY
B
24VDC/AC
V
3 4
POWER SUPPLY
24VDC
A
3 4
POWER SUPPLY
1 2
Current Output 4-20mA
Voltage Output 0-10V
RS485/Modbus RTU Output
24VDC/AC
ON
1
2
ON
1
2
ON
1
2
ON
1
2
0.5S
1S
2S
5S
S2
RS485 terminal
resistance
(inside product)
ON
1
J8
zero-reset