Calibration
24
V-04/2018
8
Calibration
Notice
The sensor unit is calibrated at the factory for operation with the
base plate oriented downward.
To operate the sensor with a different orientation of the base
plate (upward, or to the left or right), it must be calibrated in the
desired position.
Once all positions have been completely calibrated the sensor
can be used in all positions without further calibration.
Recalibration may be necessary if the sensor unit is subjected to extreme
impacts (e.g. from being dropped). Calibration is necessary if the sensor unit is
not to be operated with the base plate downward – in other words, with the
base plate upward, or to the right or left. With the exception of the base plate,
the calibration process is identical in all positions and is therefore described in
this manual only for the position with the base plate oriented downward.
In addition to the sensor unit to be calibrated, the following are needed:
Display module
Calibration cable (see illustration on page 12)
Normal XLR cables are not suitable.
Spirit level
Set square
Step 1
► Connect the sensor unit to the display module using the calibration cable.