Preventative Maintenance Guide

Dansensor CheckMate 3 (EC)

Technical Support: 01223 492087

[email protected]

Dansensor Preventative Maintenance Guides help you to care more effectively for your equipment 

during the time between planned services.

For Safety Instructions, Trouble Shooting & Operation - please refer to the User Guide.

Cleaning and Maintenance

 

In addition to the Annual Service & Calibration, it is recommended to carry out the following 

maintenance checks at the interval times stated below. 

To order consumables please contact Ann Walker on 

01223 495484 or email [email protected]

Every Day:

  

Water Trap Filter and Needle - Check and Replace (as required)

Once a Week:

 

Sample Hose – Check and Replace (as required)

During use of the device, dust, debris, and moisture may be 
sucked through the measuring system.

 

 

  

1.

The water trap filter will therefore have to be replaced 

regularly depending on the environment in which the device is 
used. 

2.

 The needle may sometimes touch the product being 

measured, or the needle may suck fluid from the packaging 
from which the measurement is taken. If this happens, it will be 
necessary to replace both the needle and the filter.

3.

 If there is any dirt or fluid in the sample hose it should be 

replaced or cleaned using dry compressed air.

To see the calibration menu, log into Access level as a 
supervisor. The standard supervisor passcode is ‘0000’. Once 
in the main menu, (Screen 1), use the down arrow key to scroll 
to the calibration menu. Push the right arrow key to enter.

The appearing calibration menu (Screen 2) holds a list of the 
various sensors in the device. For each sensor there is a table 
showing the appropriate calibration parameters. To proceed 
with the calibration, push the right arrow key.

From the next menu (Screen 3), select between the following
items: ‘Perform calibration’ or ‘Erase/enter calibration’. Select 
‘Perform calibration’ using the circle enter key.

Use the up, or down arrow key to select the calibration value 
fields for the O

 sensor (EC), then push the circle enter key.

Screen 1

Screen 2

Screen 3

Screen 4

Daily Air Calibrations

- Daily offset (20.9% O

) calibrations can be performed on the CheckMate 3 (EC) version. 

For more information please contact Technical Support. NOTE: ONLY APPLIES TO EC SENSOR VERSIONS.