Unit Entry
There are two separate access panels located at the back of the unit
4.1
Lower Access Panel Lock
The Air Medic have a Keyed Cam Lock as an optional accessory.
This lock is located on the upper middle of the lower access
panel. The key allows the unit to be locked to avoid any
unwanted entry and tampering of the unit.
4.2
Lower Access Panel
The lower access panel is the main access into the unit. this is
the panel that allows the user to access the Prefilter and the
Activated Carbon Filter. This panel is equipped with a door
microswitch. This switch stops both fans, UVC light, and
Ionizer.
This panel is equipped with two cam latches that turn to lock
in place to secure the panel, and the optional lock mentioned
above.
To open the lower access panel, unlock the cam lock, turn the
cam latches to the unlocked position and then pull up and out.
When replacing the panel, align the panel with the width of the
unit and slide it down onto the hook at the bottom. Make sure
not to damage the power cable.
4.3
Upper Access Panel
The upper access panel allows access the HEPA filter, UVC light,
and Ionizer.
This panel removed by undoing the screws located on the back
panel. There is one screw located on the bottom of the panel
only accessible with the lower panel removed. This is to ensure
all electrical devices in the unit are stopped while it is open.
Remove the panel by lifting off the guide hook and pulling
towards you.