May
2015
WorkHorse H-ADCP Operation Manual
Page 174
EAR-Controlled Technology Subject to Restrictions Contained on the Cover Page.
WB - Bandwidth Control
Purpose
Sets profiling mode 1 bandwidth (sampling rate). Narrow bandwidths allow the H-ADCP
to profile farther, but the standard deviation is increased by as much as 2.5 times.
Format
WBn
Range
n = 0 (Wide), 1 (Narrow)
Default
WB1
Recommended Setting.
The default setting for this command is recommended for most applications.
Description
See table below.
Table 28:
Bandwidth Control
Bandwidth
Sample rate
Data variance
Profiling range
0 = Wide (25%)
High
Low
Low
1 = Narrow (6.25%)
Low
High
High
WC - Low Correlation Threshold
Purpose
Sets the minimum threshold of water-track data that must meet the correlation criteria.
Format
WCnnn
Range
nnn = 0 to 255 counts
Default
WC64
Recommended Setting.
The default setting for this command is recommended for most applications.
Description
The H-ADCP uses WC to screen water-track data for the minimum acceptable correlation
requirements. The nominal (maximum) correlation depends on system frequency and
depth cell size (WS). WC sets the threshold of the correlation below, which the H-ADCP
flags the data as bad and does not average the data into the ensemble.
The default threshold for all frequencies is 64 counts. A solid target would have a correlation of
255 counts.