Table 23:
Data Stream Selections
Format
Description
PD0
Sends The real water-current data set
PD1
Sends a TRDI-defined data set that always uses the same data (except for parts of the leader data). This data set is useful
during user-software development.
PD2
Not used.
PD3
Sends Paramax-DVL ensemble output data structure.
PD4
Sends CSS-DVL output data structure (without sensor and made-good data).
PD5
Sends CSS-DVL output data structure (with sensor and made-good data).
PD6
Sends a special DVL ASCII data stream
PD7
Not used
PD8
Sends ensemble data as formatted ASCII text. A new-line character terminates each line. Two new-line characters terminate
an ensemble.
PD14
Send the H-ADCP Condensed 2D Output Format
PD16
NMEA format without the leading '$'
PD18
PD18 is the same Output Data Format as PD16, but with the leading '$' necessary to fully comply with the NMEA format.
Not all data formats can be recorded. Carefully review the
the
to set where data is sent and if the recorder is on or off.
All of TRDI’s software supports PD0 formatted data only.
The table above gives a brief overall description of the data format output. For details on the
PM - Distance Measurement Facility
Purpose
Lets you measure distance over the bottom.
Format
PM
Recommended Setting.
For TRDI use only.
Description
PM lets you use the H-ADCP to measure distances over the bottom using a dumb termi-
nal.
PS – Display System Parameters
Purpose
Sends/displays H-ADCP system configuration data.
Format
PSn
Range
n = 0, 3 (see description)
Recommended Setting.
Use as needed.
Description
See below.