UA-6 Detector
Section 4 Theory of Operation
4-5
4.2.5 Recorder Offset/Set
Lamp and Optics
U215B performs five functions:
• Reduces the signal from the noise filter from 1.53 V at
full-scale to 1.0 V.
• Provides ± 86% of scale recorder offset in conjunction
with R01.
• Ties the event mark signal to the recorder signal.
• Ties the inject mark signal to the recorder signal.
• When the sensitivity range switch is set on SET LAMP
& OPTICS and the optical unit is properly balanced,
U215B works in conjunction with Q204, U213A and
S05B to offset the recorder signal to 0% of scale.
4.2.6 Peak Separation
U215D offsets and amplifies the noise filter signal for the peak
separator. The peak separator, (U401 and U404), outputs a signal
that is proportional to the edge slope, or time derivative, of a
detected peak. A fast slope causes the output of U404B to be
greater than a slow slope. The PEAK DURATION switch (S506),
compensates for this fact by making the detector more sensitive
for slow peaks. TP402 will give a reading between 0 and 8.5 V for
negative peak edges and a reading between 0 and -8.5 V for pos-
itive peak edges.
Operational amplifier U404D switches from -13 V to +13 V when
a negative edge is detected. Operational amplifier U404C
switches from -13 V to +13 V when a positive edge is detected.
Capacitor C442 or C424 makes a pulse whenever U404D or
U404C is turned on. These pulses are used to index a fraction col-
lector when a new peak is detected or an old peak is finished.
U406A and Q414 produce a peak duration signal (TTL low to
P107, J217 during a peak).
4.2.7 Injection and Event
Mark
A positive spike can be placed on the chart recorder by shorting
pin 4 to pin 9 of J217, the INJECTION MARK input. This action
will cause C232 to pass a pulse to operational amplifier U215C
and then on to the recorder through the optical unit connector.
(Pins 14 and 13 are shorted when the optical unit is plugged in.)
The negative event mark spike works the same way when pins 3
and 9 of the FRACTION COLLECTOR connector are shorted
together. Here the pulse is made by C228 and operational
amplifier U215A.
4.2.8 Lamp Brightness
Regulator
Located in the left center portion of the schematic, this circuit is
only active when UA-6 is used with a Type 11 optical unit. When
active, it performs five functions:
• 10 minute warm-up cycle.
• Limits lamp current to a maximum of 270 mA.
• Automatically adjusts lamp current to a pre-selected
value. (Value is selected by user, S401.)
• Monitors lamp current.
• Maintains lamp brightness.
Содержание UA-6
Страница 2: ......
Страница 4: ......
Страница 12: ...x ...
Страница 16: ...Isco UA 6 Table of Contents xiv ...
Страница 32: ...UA 6 Detector Section 2 Preparation for Use 2 8 ...
Страница 41: ...UA 6 Detector Section 3 Operation 3 9 Figure 3 6 Ferrule Application Information ...
Страница 47: ...UA 6 Detector Section 3 Operation 3 15 Figure 3 9 Absorbance Curves of Four Commonly Used Solvents ...
Страница 52: ...UA 6 Detector Section 3 Operation 3 20 Figure 3 14 Problem Analysis Drift and Noise ...
Страница 54: ...UA 6 Detector Section 3 Operation 3 22 ...
Страница 65: ...UA 6 Detector Section 4 Theory of Operation 4 11 Figure 4 5 Type 11 Absorbance Detector Block Diagram ...
Страница 66: ...UA 6 Detector Section 4 Theory of Operation 4 12 ...
Страница 90: ...UA 6 Detector Appendix A Replacement Parts A 2 ...
Страница 91: ...UA 6 Detector Appendix A Replacement Parts A 3 ...
Страница 92: ...UA 6 Detector Appendix A Replacement Parts A 4 MAIN CIRCUIT BOARD ...
Страница 93: ...UA 6 Detector Appendix A Replacement Parts A 5 ...
Страница 94: ...UA 6 Detector Appendix A Replacement Parts A 6 ...
Страница 95: ...UA 6 Detector Appendix A Replacement Parts A 7 ...
Страница 96: ...UA 6 Detector Appendix A Replacement Parts A 8 ...
Страница 97: ...UA 6 Detector Appendix A Replacement Parts A 9 ...
Страница 98: ...UA 6 Detector Appendix A Replacement Parts A 10 ...
Страница 99: ...UA 6 Detector Appendix A Replacement Parts A 11 ...
Страница 100: ...UA 6 Detector Appendix A Replacement Parts A 12 ...
Страница 101: ...UA 6 Detector Appendix A Replacement Parts A 13 ...
Страница 102: ...UA 6 Detector Appendix A Replacement Parts A 14 ...
Страница 103: ...UA 6 Detector Appendix A Replacement Parts A 15 ...
Страница 104: ...UA 6 Detector Appendix A Replacement Parts A 16 ...
Страница 105: ...UA 6 Detector Appendix A Replacement Parts A 17 ...
Страница 106: ...UA 6 Detector Appendix A Replacement Parts A 18 This page intentionally left blank ...
Страница 107: ...UA 6 Detector Appendix A Replacement Parts A 19 ...
Страница 108: ...UA 6 Detector Appendix A Replacement Parts A 20 ...
Страница 109: ...UA 6 Detector Appendix A Replacement Parts A 21 Drawing modified 7 6 11 to remove obsolete flow cells ...
Страница 110: ...UA 6 Detector Appendix A Replacement Parts A 22 ...
Страница 111: ...UA 6 Detector Appendix A Replacement Parts A 23 ...
Страница 112: ...UA 6 Detector Appendix A Replacement Parts A 24 ...
Страница 114: ...UA 6 Detector Index Index 2 Lamp Replacement 5 11 Theory of Operation 4 8 Type 12 Fuses 2 4 ...
Страница 116: ......
Страница 118: ......
Страница 120: ...Warranty ...
Страница 122: ......