08354B DCN7877
Teledyne API T640 PM Mass Monitor
44
flash drive is the same as or older than the current firmware of the
instrument, the Install button will not be enabled.
5. Press the Install button, and note the messages in the Status field at
the bottom of the page. Use the Cancel button if necessary.
6. When complete, as indicated in the Status field, press the Done button,
which replaces the Cancel button, and remove the flash drive.
7. Power off and restart the instrument to complete the new firmware
installation.
TRANSFERRING CONFIGURATION TO OTHER INSTRUMENTS
Once an instrument is configured, the same configuration can be copied to other
instruments of the same Model. This encompasses essentially anything the user can
configure and does not apply to instrument-specific settings such as those that are
configured at the factory for calibration.
Figure 3-8:
Configuration Transfer
1. In the source instrument, navigate to the Home>Utilities>USB
Utilities>General page.
2. Insert a flash drive into either of the two front panel USB ports.
3. When the Status field indicates that the USB drive has been detected,
press the “Download Configuration from Instrument” Start button.
4. When the Status field indicates that the download is complete, remove
the flash drive.
5. In the target instrument, navigate to the Home>Utilities>USB
Utilities>General page.
6. Insert a flash drive into either of the two front panel USB ports.
7. When the Status field indicates that the USB drive has been detected,
press the “Upload Configuration to Instrument” Start button.
When the Status field indicates that the upload is complete, remove the flash drive.