08354B DCN7877
Teledyne API T640 PM Mass Monitor
43
Manually Prompted Updates
The instrument must be connected to a network that is connected to the Internet. In
the Setup>Instrument>Remote Update menu press the Check for Updates button.
If the Status field shows that an update is available, it can be downloaded through
this page.
Figure 3-6:
Remote Update Page
Manual Reload/Update Procedures
To reload or update firmware, first contact Technical Support to obtain the
applicable file(s): [email protected] / 800-324-5190.
1. Follow Technical Support’s instructions for copying the firmware files to
a flash drive.
2. Navigate to the Utilities>USB Utilities>General menu.
Figure 3-7: Manual Update Page (and other utilities)
3. Insert a flash drive into a front panel USB port and wait for the Status
field to indicate that the drive has been detected.
4. In the Update Firmware field, press the Check button for the instrument
to determine whether the firmware on the flash drive is more recent than
what is currently installed. Once it’s been determined that the firmware
is new, the Install button will be enabled; if the firmware version on the