Teledyne API Model T360/T360M Operation Manual
Principles of Operation
211
7.4.4.2. SIGNAL SYNCHRONIZATION AND DEMODULATION
The signal emitted by the IR photo-detector goes through several stages of
amplification before it can be accurately demodulated. The first is a pre-
amplification stage that raises the signal to levels readable by the rest of the
synch/demod board circuitry. The second is a variable amplification stage that is
adjusted at the factory to compensate for performance variations of mirrors,
detectors, and other components of the optical bench from instrument to
instrument.
The workhorses of the sync/demod board are the four sample-and-hold circuits
that capture various voltage levels found in the amplified detector signal needed
to determine the value of
CO2 MEAS
and
CO2 REF
. They are activated by
logic signals under the control of a compact programmable logic device (PLD),
which in turn responds to the output of the Segment Sensor and M/R Sensor
depicted in Figure 7-14.
The four sample and hold circuits are:
ACTIVE WHEN:
DESIGNATION
IR BEAM PASSING THROUGH
SEGMENT SENSOR PULSE IS:
Measure Gate
MEASUREMENT cell of GFC Wheel
HIGH
Measure Dark Gate
MEASUREMENT cell of GFC Wheel
LOW
Reference Gate
REFERENCE cell of GFC Wheel
HIGH
Reference Dark Gate
REFERENCE cell of GFC Wheel
LOW
Timing for activating the Sample and Hold circuits is provided by a phase lock
loop circuit (PLL). Using the segment sensor output as a reference signal the
PLL generates clock signal at ten times that frequency. This faster clock signal is
used by the PLD to make the sample and hold circuits capture the signal during
the center portions of the detected waveform, ignore the rising and falling edges
of the detector signal.
Sample & Hold
Active
Detector
Output
Sample & Hold
Inactive
Figure 7-16:
Sample & Hold Timing
07272B DCN6552
Содержание T360
Страница 2: ......
Страница 4: ...Model T360 T360M Operation Manual Teledyne API ii This page intentionally left blank 07272B DCN6552 ...
Страница 8: ...Model T360 T360M Operation Manual Teledyne API vi This page intentionally left blank 07272B DCN6552 ...
Страница 10: ...Model T360 T360M Operation Manual Teledyne API viii This page intentionally left blank 07272B DCN6552 ...
Страница 18: ...Teledyne API Model T360 T360M Operation Manual xvi This page intentionally left blank 07272B DCN6552 ...
Страница 188: ...This page intentionally left blank 07272B DCN6552 ...
Страница 264: ...07272B DCN6552 ...
Страница 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Страница 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Страница 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Страница 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Страница 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Страница 333: ...07272B DCN6552 D 7 ...
Страница 338: ...07272B DCN6552 D 12 ...