Operating Instructions
Teledyne API Model T360/T360M Operation Manual
100
4.8.5. Auto Range Mode (AUTO)
In
AUTO
range mode, the analyzer automatically switches the reporting range
between two user-defined ranges (low and high). The unit will switch from low
range to high range when the CO
2
concentration exceeds 98% of the low range
span. The unit will return from high range back to low range once both the CO
2
concentration falls below 75% of the low range span.
In
AUTO
Range mode the instrument reports the same data in the same range on
both the
A1
and
A2
outputs and automatically switches both outputs between
ranges as described above. Also, the
RANGE
test function displayed on the front
panel will automatically switch to show which range is in effect.
The high/low range status is also reported through the external, digital status
outputs (Section 4.15.1).
To set individual ranges press the following menu button sequence.
SAMPLE* RANGE = 500.000 PPM CO2 =X.XXX
< TST TST > CAL
SETUP
SETUP X.X
PRIMARY SETUP MENU
CFG DAS
RNGE
PASS CLK MORE
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET UNIT
EXIT
SETUP X.X
RANGE MODE:
SNGL
SNGL DUAL
AUTO
ENTR EXIT
EXIT
x 2 returns
to the main
SAMPLE
display
SETUP X.X
RANGE MODE:
AUTO
SNGL
DUAL AUTO
ENTR
EXIT
SETUP X.X
RANGE CONTROL MENU
MODE
SET
UNIT
EXIT
SETUP X.X
HIGH RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
SETUP X.X
LOW RANGE:
500.0 Conc
0 0 5 0 0 .0
ENTR
EXIT
Toggle the numeral
buttons to set the
LOW
and
HIGH
range value.
ENTR accepts the
new setting, EXIT
ignores the new
setting.
NOTE
In AUTO range mode the LOW and HIGH ranges have separate slopes and offsets for computing
CO2 concentration. The two ranges must be independently calibrated.
NOTE
Avoid accidentally setting the low range of the instrument with a higher span limit tha The high
range. This will cause the unit to stay in the low reporting range perpetually and defeat the
function of the AUTO range mode.
07272B DCN6552
Содержание T360
Страница 2: ......
Страница 4: ...Model T360 T360M Operation Manual Teledyne API ii This page intentionally left blank 07272B DCN6552 ...
Страница 8: ...Model T360 T360M Operation Manual Teledyne API vi This page intentionally left blank 07272B DCN6552 ...
Страница 10: ...Model T360 T360M Operation Manual Teledyne API viii This page intentionally left blank 07272B DCN6552 ...
Страница 18: ...Teledyne API Model T360 T360M Operation Manual xvi This page intentionally left blank 07272B DCN6552 ...
Страница 188: ...This page intentionally left blank 07272B DCN6552 ...
Страница 264: ...07272B DCN6552 ...
Страница 320: ...This page intentionally left blank B 2 07272B DCN6552 ...
Страница 324: ...This page intentionally left blank B 6 07272B DCN6552 ...
Страница 327: ...APPENDIX D Wire List and Electronic Schematics D 1 07272B DCN6552 ...
Страница 328: ...This page intentionally left blank D 2 07272B DCN6552 ...
Страница 332: ...This page intentionally left blank 07272B DCN6552 D 6 ...
Страница 333: ...07272B DCN6552 D 7 ...
Страница 338: ...07272B DCN6552 D 12 ...