Getting Started
Teledyne API – Model T300/T300M CO Analyzer
70
3.3.2.7.
PNEUMATIC CONNECTIONS FOR ZERO SCRUBBER/PRESSURIZED SPAN OPTION
Source of
SAMPLE GAS
Removed during
calibration
VENT
Calibrated
CO Gas
at span gas
concentration
(Adjust to 30
psig)
VENT
here if input
is pressurized
PRESSURE SPAN
ZERO AIR
SAMPLE
EXHAUST
VENT SPAN
Instrument
Chassis
Figure 3-25:
Pneumatic Connections – Zero Scrubber/Pressurized Span Calibration Valves (Opt 50E)
S
AMPLE
G
AS
S
OURCE
Attach a sample inlet line to the sample inlet port. The SAMPLE input line should not
be more than 2 meters long.
•
Maximum pressure of any gas at the sample inlet should not exceed 1.5 in-hg above
ambient pressure and ideally should equal ambient atmospheric pressure.
•
In applications where the sample gas is received from a pressurized manifold, a
vent must be placed on the sample gas before it enters the analyzer.
C
ALIBRATION
G
AS
S
OURCES
S
PAN
G
AS
:
•
Attach a gas line from the pressurized source of calibration gas (e.g. a bottle of
NIST-SRM gas) to the span inlet.
•
Span gas can by generated by a T700 Dynamic Dilution Calibrator.
Z
ERO
A
IR
:
•
Zero air is supplied internally via a zero air scrubber that draws ambient air through
the ZERO AIR inlet.
I
NPUT
G
AS
V
ENTING
The zero air supply and sample gas line MUST be vented in order to ensure that the
gases input do not exceed the maximum inlet pressure of the analyzer as well as to
prevent back diffusion and pressure effects. These vents should be:
•
At least 0.2m long;
•
No more than 2m long and;
•
Vented outside the shelter or immediate area surrounding the instrument.
A similar vent line should be connected to the VENT SPAN outlet on the back of the
analyzer.
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...