
31
2.
SPECIFICATIONS AND APPROVALS
This section presents specifications for the T300/T300M analyzer and for its second gas
sensor options, EPA equivalency designation, and compliance statements.
2.1.
SPECIFICATIONS
Table 2-1: T300/T300M Basic Unit Specifications
Parameter
Specification
Ranges
Min: 0-1 ppm Full scale
Max: 0-1,000 ppm Full scale (selectable, dual ranges and auto ranging supported)
Measurement Units
T300: ppb, ppm, µg/m
3
, mg/m
3
(user selectable)
T300M: ppm, mg/m
3
(user selectable)
Zero Noise
1
T300: < 0.02 ppm RMS
T300
M: ≤ 0.1 ppm RMS
Span Noise
1
T300:<0.5% of rdg RMS over 5ppm
3
T300M:>0.5% of rdg RMS over 20ppm
Lower Detectable Limit
1
T300: < 0.04 ppm
T300M: 0.2 ppm
Zero Drift (24 hours)
2
T300: < 0.1 ppm
T300M: <0.5 ppm
Span Drift (24 hours)
2
T300: < 0.5% of reading
T300M: 0.5ppm
Lag Time
1
10 seconds
Rise/Fall Time
1
<60 seconds to 95%
Linearity
T300: 1% of full scale
5
;
T300M: 0 - 3000 ppm: 1% full scale;
3000 - 5000 ppm: 2% full scale
Precision
T300: The greater of 0.5% of reading or 0.2ppm;
T300M: The greater of 1.0% of reading or 1ppm
Sample Flow Rate
800 cm
3
/min. ±10%
(O
2
Sensor option adds 120 cm³/min to total flow when installed)
AC Power
100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges
All Outputs: 10V, 5V, 1V, 0.1V (selectable)
Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over-range
Analog Output Resolution
1 part in 4096 of selected full-scale voltage
Recorder Offset
±10%
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs
6 opto-isolated digital control inputs (2 defined, 4 spare)
4 user configurable analog outputs
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...