Theory of Operation
Teledyne API – Model T300/T300M CO Analyzer
300
IR
Source
Photo-Detector
M
R
GFC Wheel
IR unaffected by N
2
in Measurement Cell
IR is affected by CO in Reference Cell
Δ H
Figure 13-3:
Measurement Fundamentals with GFC Wheel
As the GFC Wheel spins, the IR light alternately passes through the two cavities. When
the beam is exposed to the reference cell, the CO in the gas filter wheel strips the beam
of most of the IR at 4.7
μm. When the light beam is exposed to the measurement cell,
the N
2
in the filter wheel does not absorb IR light. This causes a fluctuation in the
intensity of the IR light striking the photo-detector which results in the output of the
detector resembling a square wave.
13.2.1.2.
THE MEASURE REFERENCE RATIO
The T300/T300M determines the amount of CO in the sample chamber by computing
the ratio between the peak of the measurement pulse (
CO MEAS
) and the peak of the
reference pulse (
CO REF
).
If no gases exist in the sample chamber that absorb light at 4.7
μm, the high
concentration of CO in the gas mixture of the reference cell will attenuate the intensity
of the IR beam by 60% giving a M/R ratio of approximately 2.4:1.
Adding CO to the sample chamber causes the peaks corresponding to both cells to be
attenuated by a further percentage. Since the intensity of the light passing through the
measurement cell is greater, the effect of this additional attenuation is greater. This
causes
CO MEAS
to be more sensitive to the presence of CO in the sample chamber
than
CO REF
and the ratio between them (M/R) to move closer to 1:1 as the
concentration of CO in the sample chamber increases.
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...