Teledyne API – Model T300/T300M CO Analyzer
Troubleshooting and Service
285
SAMPLE
RANGE=50.0 PPM CO= XXXX
<TST
TST> CAL
SETUP
DIAG
SIGNAL I/O
PREV
NEXT
ENTR
EXIT
DIAG AOUT
ANALOG OUTPUT
[20%]
EXIT
SETUP X.X
ENTER PASSWORD
8
1
8
ENTR
EXIT
DIAG AOUT
ANALOG OUTPUT
20%
EXIT
•
Pressing the button under “0%” pause
the test. Brackets will appear around the
value: EXAMPLE: [20%]
•
Pressing the same button again will
resume the test.
Performs analog output step
test 0% to 100%
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
12.5.8.3.
ANALOG OUTPUTS: CURRENT LOOP
To verify that the analog outputs with the optional current mode output are working
properly, connect a 250 ohm resistor across the outputs and use a voltmeter to measure
the output as described in Section 5.9.3.6 and then perform an analog output step test as
described in Section 12.5.8.2.
For each step the output should be within 1% of the nominal value listed in the table
below.
Table 12-11:
Analog Output Test Function - Nominal Values Voltage Outputs
OUTPUT RANGE
2 -20
4 -20
NOMINAL OUTPUT VALUES
STEP
%
CURRENT
V(250 OHMS)
CURRENT
V(250 OHMS)
1
0
2 mA
0.5V
4
1
2
20
5.6
1.4
7.2
1.8
3
40
9.2
2.3
10.4
2.6
4
60
12.8
3.2
13.6
3.4
5
80
16.4
4.1
16.8
4.2
6
100
20
5
20
5
06864B DCN6314
Содержание T300
Страница 2: ......
Страница 4: ...Teledyne API Model T300 T300M CO Analyzer ii This page intentionally left blank 06864B DCN6314 ...
Страница 8: ...Teledyne API Model T300 T300M CO Analyzer vi This page intentionally left blank 06864B DCN6314 ...
Страница 12: ...Teledyne API Model T300 T300M CO Analyzer x This page intentionally left blank 06864B DCN6314 ...
Страница 25: ...23 PART I GENERAL INFORMATION 06864B DCN6314 ...
Страница 26: ...24 06864B DCN6314 ...
Страница 88: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 86 This page intentionally left blank 06864B DCN6314 ...
Страница 89: ...87 PART II OPERATING INSTRUCTIONS 06864B DCN6314 ...
Страница 90: ...Getting Started Teledyne API Model T300 T300M CO Analyzer 88 06864B DCN6314 ...
Страница 144: ...Setup Menu Teledyne API Model T300 T300M CO Analyzer 142 This page intentionally left blank 06864B DCN6314 ...
Страница 198: ...Remote Operation Teledyne API Model T300 T300M CO Analyzer 196 This page intentionally left blank 06864B DCN6314 ...
Страница 243: ...241 PART III TECHNICAL INFORMATION 06864B DCN6314 ...
Страница 244: ...06864B DCN6314 ...
Страница 245: ...243 06864B DCN6314 ...
Страница 246: ...06864B DCN6314 ...
Страница 346: ...INDEX Teledyne API Model T300 T300M CO Analyzer 344 This page intentionally left blank 06864B DCN6314 ...
Страница 386: ...This page intentionally left blank B 2 06864B DCN6314 ...
Страница 395: ...APPENDIX D Wire List and Electronic Schematics 06864B DCN6314 D 1 ...
Страница 396: ...This page intentionally left blank D 2 06864B DCN6314 ...
Страница 400: ...This page intentionally left blank D 6 06864B DCN6314 ...
Страница 401: ...06864B DCN6314 D 7 ...